Hermetic Circular and Bulkhead Connectors

When electrical circuits cross a sealed housing wall, first decide where the gas barrier sits and whether installers need to disconnect the electrical connection. A hermetic circular receptacle can combine a removable plug interface with sealed conductors. A bulkhead feedthrough can carry those conductors through the wall and end in pins or terminals for permanent wiring. The housing joint also needs its own seal definition and acceptance criteria. For a drawing-based project, send MOOSEN the wall section, contact layout and leak-test requirement so we can review the requested component and its interfaces.

Quality inspection workspace at MOOSEN

Circular connectors, bulkhead receptacles and feedthroughs

Circular describes the mating interface. The outer mounting flange may still be square. A circular connector drawing should identify the connector series or mating geometry, shell size, coupling, key position and contact arrangement. Shell size and contact count describe different features. A supplier needs both to identify which insert and mating half fit.

Bulkhead describes installation across a panel, partition or enclosure wall. A bulkhead assembly can contain a circular or rectangular receptacle, or it can be a feedthrough with permanent terminations. Mounting and sealing deserve separate drawing callouts. A nut or bolted flange provides mechanical retention. The drawing should show how the resulting joint seals against the panel. Published circular connector catalogs illustrate distinct solder-mount, weld-mount, flange-mount and jam-nut receptacles. See the Amphenol Aerospace hermetic solutions catalog.

Receptacle identifies the connector half that receives the mating plug. A panel-mounted receptacle can use pin or socket contacts, so specify the contact style and drawing view. Feedthrough describes the function of carrying a conductor through a sealed barrier. A hermetic receptacle can serve as a feedthrough. Bare-ended feedthroughs can also use solder tails, wire leads or other terminations designed into the assembly. See the Amphenol Aerospace connector reference guide.

FormatWhen to evaluate itWhat the drawing should settle
Hermetic circular receptacleA serviceable cable needs to mate with a round interface at a sealed boundary.Mating dimensions, shell size, coupling, keying, insert arrangement and rear termination.
Hermetic bulkhead connectorAn electrical connection crosses an enclosure wall and the installation must preserve the barrier.Panel geometry, mounting side, retaining hardware or permanent joint, and leakage after installation.
Hermetic feedthrough or headerConductors pass through the wall and connect to an internal circuit, wire or contact assembly.Conductor geometry, insulation, pin position, housing interface and the component test boundary.
Rectangular connector or single-seal terminalThe enclosure calls for a rectangular mating face or needs one independently located conductor.Opening and contact map, or conductor dimensions and the terminal-to-housing joint.
Connector and feedthrough formats for sealed housings

Specify the connector pin count and contact layout

Start with a circuit list. Give each position a number, function, termination and electrical requirement. Mark intended spares, empty positions and any provision for a separate ground or shell bond. Identify where the pins or sockets appear, and label each view as the mating face or wiring side. For a keyed interface, dimension the pin pattern from the key or another defined datum. Include the mating connector part number and drawing if the plug is already selected.

Two connectors with the same pin count can use different spacing, contact sizes or key positions. A pin count also leaves open how much room the back of the component needs for a solder joint, a wire bend or a board connection. Show pin diameter and position tolerance, exposed length and termination on both sides. Include limits for relevant electrical checks such as continuity, insulation resistance and dielectric withstand, with the test points identified. If the assembly mixes power and signals, provide the voltage and current requirements for each applicable circuit so the contact layout can be evaluated with the application in view.

Choose the shell, seal construction and mounting interface

On a circular receptacle, the shell locates and protects the mating interface. In a feedthrough assembly, the body also provides the interface to the host housing. Specify the shell or body alloy and finish, mating dimensions, panel material and the installed envelope. Include the clearance needed to turn the coupling, route wiring and reach fasteners. If the finish has to leave a joining surface exposed, show that area on the drawing.

A glass-to-metal seal uses glass to electrically isolate the conductors and form the barrier between the pins and metal body. Matched and compression seal designs use different relationships between the thermal expansion of the glass and metals. In a ceramic-to-metal design, an alumina insulator and brazed joints are examples of materials and construction used to build a sealed feedthrough. The suitable material stack depends on insulation requirements, geometry, temperature exposure and how the assembly will be joined into the housing. See SCHOTT’s glass-to-metal sealing overview and CeramTec’s ceramic-to-metal feedthroughs.

Show the intended shell-to-wall joint in section. A weld, a soldered joint and a retained fitting each create a different interface for inspection. For permanent assembly, provide the proposed joining process and any relevant temperature or access constraints. For removable mounting, define the sealing element, its seating surface and the panel thickness range. When the design uses a separate gasket at a flange, identify that joint explicitly so the component leak test and enclosure leak test each cover the intended boundary.

Write a measurable helium leak-test requirement

Give the maximum allowed leak rate with units and the gas basis for reporting. Specify the test method or referenced procedure, tracer gas and concentration, pressure on each side or the applicable differential, and the direction of leakage being evaluated. Include temperature, timing and conditioning when the procedure calls for them. A helium vacuum test and a helium sniffer test use different setups. State how the part is fixtured and whether it is tested unmated, mated or already mounted to the housing. See INFICON’s leak-test method comparison.

A supplier may measure leakage through the standalone component in a fixture, while the OEM measures the finished enclosure after the component is installed. Name the surfaces and seals included at each stage. If assembly applies heat to the joint or the qualification plan calls for thermal cycling or another exposure, agree on the order of tests and the acceptance points before quoting. Where the project calls for both fine and gross leak checks, list the procedure and acceptance criteria for each. This lets quality teams compare reports against the same written requirements and identify which joint a test result represents.

Match the component format to the housing and service plan

Start with who needs access after assembly. A serviceable instrument with an external cable may benefit from a circular receptacle. Inside the enclosure, a custom feedthrough may terminate directly onto a board or an internal harness. A wide, shallow opening can also justify evaluating a hermetic rectangular connector. If the housing only needs one conductor at a specific location, review the geometry and joining interface of a single-seal terminal.

For medical OEM projects, give the installation drawing, interfaces and project-specific documentation requirements early. A component may be one part of a larger sealed enclosure, alongside contacts, insulating parts and the final housing assembly. MOOSEN’s medical hermetic feedthrough components and implantable connection components pages describe those component-level options. Use the project drawing to identify exactly which part you want quoted and which assembly steps the OEM will perform.

Check drawing compatibility when changing suppliers

For an existing design, identify the mating plug, current receptacle or feedthrough, and every interface the replacement must fit. Include the original part number with its full suffix, an approved drawing or permitted interface dimensions, and the configuration used in your equipment. A shared series name or matching diameter leaves key rotation, pin projection, panel fit and rear clearance to verify. If a particular feature must stay interchangeable across lots, specify its tolerance and where your inspection fixture references it.

For a new design, mark the dimensions that are fixed by the housing or the mating plug and identify the dimensions that are open for engineering review. Share the assembly sequence where it affects access. You may be able to test the feedthrough before closing the housing and have access only to the external contacts afterward. Agree on how to assess the first samples against the drawing, including how to capture fit, electrical checks and leakage at the required stage.

What to include in a drawing-based RFQ

Send the current drawing revision and specify whether you need a complete mating connector assembly, a bulkhead-mounted receptacle or a sealed component for incorporation into your own equipment. A section of the housing and the mating or termination drawing usually answer more sourcing questions than an envelope sketch alone.

  • Mechanical interface. Include an assembly section, panel material and thickness, opening geometry and tolerances, datums, minimum clearance, mounting side and retention or joining details.
  • Electrical interface. Include contact count and map, pin or socket geometry, key orientation, mating part reference, termination on both sides and electrical acceptance limits.
  • Materials and environment. List metal and insulator restrictions, plating or finish, pressure or vacuum conditions, operating temperature and exposures from downstream assembly or service.
  • Inspection and records. Call out critical dimensions, visual criteria, leak-test method and limits, electrical checks, sample size or unit-level inspection, and the reports required with shipment.
  • Commercial scope. Identify prototype quantity, expected production volume, target dates and the party responsible for the panel joint, mating plug and enclosure-level verification.

Align the inspection plan with the drawing revision. If your incoming inspection references particular datums or dimensions, mark those features for the supplier so the reported measurements can be compared. If lot traceability, first-article records or actual test readings are needed, state the required format and how records should link to the shipment. See MOOSEN’s quality information and include any additional reporting needs with the RFQ.

Request a review of your circular or bulkhead interface

Send MOOSEN your drawing, contact layout and leak-test conditions for a documented engineering review. Include any interface that must fit an existing plug and any mounting or test requirements still being finalized. We can review the requested part against the drawing and discuss component scope, interfaces and inspection records before quoting.