Type S Thermocouple Wire

A Type S thermocouple uses platinum with a nominal 10% rhodium by mass for its positive leg and high-purity platinum for its negative leg. You may see the pair written as PtRh10/Pt and the individual legs identified as SP and SN. The two wires form the measuring junction of a noble-metal thermocouple. A quotation for Type S wire should say whether it covers those sensing legs, a cable for connecting an existing probe to an instrument or an assembled sensor. Each serves a different part of the measurement circuit. See NIST Monograph 175.

Close-up of wire reels at MOOSEN

Type S wire composition and polarity

Wire or pairPositive sensing legNegative sensing leg
Type SPlatinum with nominally 10% rhodium (SP)High-purity platinum (SN)
Type RPlatinum with nominally 13% rhodium (RP)High-purity platinum (RN)
Type BPlatinum with nominally 30% rhodium (BP)Platinum with nominally 6% rhodium (BN)
Platinum-iridium wireDetermined by the component specificationDetermined by the component specification
How Type S compares with Type R, Type B and platinum-iridium wire

Rhodium (Rh) and iridium (Ir) are distinct elements. PtIr10 and PtRh10 therefore specify different alloys. Where a drawing names platinum-iridium for an electrode, lead or another component, carry that material specification through procurement. Where the drawing names a Type S measuring junction, specify SP and SN and the required thermoelectric tolerance. If both materials appear on one drawing, give each a separate identifier and state its function at the assembly level. See ASTM E1159 for platinum-rhodium and platinum thermocouple materials.

Type R and Type S share a nominal high-purity platinum negative leg, which makes the positive leg particularly important when identifying loose wires. Keep the supplier’s tags through cutting, cleaning and assembly, and label the legs in the traveler before making the junction. At the terminals, connect SP to the Type S positive input or matching positive cable conductor and SN to the corresponding negative side. Where there is a jacketed cable, verify the governing convention and the cable marking. The thermocouple wire color code chart explains how to identify a convention separately from the printed cable grade.

The Type S reference function and the actual probe

The NIST ITS-90 reference tables describe the nominal electromotive force (EMF) of a Type S pair at a given measuring-junction temperature with a reference junction at 0 °C. For instance, the nominal output is 9.587 mV at 1,000 °C and 11.951 mV at 1,200 °C. These values help verify the instrument setting or interpret a voltage measured using a known reference. The unequal voltage increments per degree across the full curve are the reason a Type S instrument uses a table or polynomial when it converts the reading. See the Omega Type S reference table and IEC 60584-1.

The NIST Type S reference function spans −50 to 1,768 °C. Set the operating range for a real assembly using the supplier’s rating for its wire diameter, insulation, protective tube or sheath, and intended atmosphere. Include the duration of a high-temperature hold and the frequency of excursions. If an installation cycles between room temperature and process temperature, give the expected cycle profile as well as the peak. Compare recurring operation and short excursions separately when reviewing a probe datasheet. See Heraeus platinum thermocouple wire guidance.

A laboratory comparison and a furnace installation can use the same reference function with different assembly requirements. For example, a laboratory may need calibrated readings at selected points with a stated uncertainty. A furnace installation may need a protective tube compatible with the atmosphere and enough insertion depth to limit heat conduction along the stem. In both cases, the wire specification is one input to a measurement design. The installation establishes how closely the junction tracks the temperature of the material or space being measured. See NIST thermocouple calibration services.

Thermocouple-grade wire and Type S compensating cable

ItemCircuit positionSpecification to check
Thermocouple-grade SP and SNMeasuring junction and sensing legs to their designed termination.Thermoelectric tolerance, leg identification, diameter, material or calibration records.
Type S compensating cableProbe termination to the compensated input, where permitted by the design.Exact designation such as IEC SCA or SCB, tolerance, insulation and temperature interval.
Instrument output wiringDownstream of a transmitter that has already converted the thermocouple signal.Wiring for the specified transmitter output.
Which conductors belong at the junction and on the lead run

Extension-grade conductors for thermocouples use the same nominal alloy system as their matching sensing pair and have a defined tolerance across an extension temperature interval. Compensating cable uses other alloys to approximate the thermoelectric behavior of the corresponding pair over a defined interval. Some catalogs use “Type S extension wire” broadly for an interconnection product. Request the exact designation and conductor description if a listing uses that phrase. IEC SCA and SCB designate Type S compensating cables, and their specified temperature intervals and initial thermoelectric deviations differ. See IEC 60584-3 and the Pentronic cable and connector guide.

Check the maximum temperature at the terminal head where the platinum-based wires join the compensating cable. That point may warm during a long furnace hold, especially if the head sits close to the hot wall or is inside an enclosure. Its temperature and the cable route need to stay within the limits applicable to the selected grade and insulation. For an existing installation, measure the head temperature during representative operation before qualifying a replacement cable. A higher jacket rating addresses insulation survival, while the compensating alloy pair has its own range of thermoelectric accuracy.

Reference junction, instrument and calibration

With cold-junction compensation enabled, the instrument needs the temperature of the terminals where the Type S circuit transitions to its copper input conductors. It converts the terminal temperature into the equivalent Type S EMF, adds that value to the measured voltage, then converts the total to temperature using the Type S curve. Configure the input for Type S and keep both terminals at a common temperature near the compensation sensor. If the design uses a separate reference setup, follow the readout manufacturer’s procedure for that input. See the Analog Devices thermocouple measurement guide.

Distinguish the initial tolerance of the material from the calibration of a supplied probe. A wire lot report can establish conformance for material tested against a specification. A calibration report can record the deviations of a particular sensor at agreed comparison temperatures. Ask which item is identified on the report, what reference was used and what uncertainty applies. In critical use, look beyond the initial calibration and establish when the installed sensor needs to be checked or replaced.

Changes in a wire’s thermoelectric properties can matter when they occur where the wire lies in a temperature gradient. A sensor exposed to repeated high-temperature service may therefore behave differently after it is moved to another depth or another furnace, even if its junction appears intact. ASTM E220 describes comparison calibration and identifies potential inhomogeneity as a limitation for used thermocouples. For measurements with a defined uncertainty budget, include the cable, readout, reference-junction compensation and installation alongside the sensor results. See ASTM E220 comparison calibration.

As an example, an input check can verify the transmitter configuration and the continuity of the compensating cable before a furnace run. The installed temperature measurement still depends on how the probe is positioned. When comparing a new probe with a reference, place the two junctions where they sample comparable conditions, allow the furnace to stabilize, and record the insertion depths. A difference between readings can reflect a genuine temperature gradient. The comparison conditions belong in the verification record alongside the instrument settings and serial numbers.

If the application has a narrow allowable error, discuss the expected sensor output at the operating temperature with the readout supplier. The input resolution, electrical noise and cold-junction compensation contribute to the uncertainty of a measurement made with a millivolt-level signal. A report for the wire alone cannot characterize those contributions once the sensor is connected and installed.

Information to include with a Type S inquiry

  • Item and function. State whether the inquiry covers an SP or SN leg, a matched sensing pair, Type S compensating cable or a finished probe. List separate items when the assembly needs both sensing wire and cable.
  • Thermoelectric acceptance. Give the applicable standard and tolerance, the temperature interval, and any calibration points or maximum uncertainty required for the finished instrument or sensor.
  • Wire and assembly dimensions. Include the diameter and tolerance of each leg, lengths, bend or routing constraints, hot-junction geometry, and connector or terminal details. For a protected probe, specify the available outside diameter and insertion depth.
  • Exposure and duty cycle. Record the normal measurement range, maximum excursions and their duration, atmosphere, and estimated head and cabinet temperatures. Describe any repeatable temperature cycles or maintenance intervals relevant to qualification.
  • Packaging and verification. Specify leg identification, cut lengths or spools, sample and production quantities, and the documents required at receipt. State whether traceability needs to connect each probe or length back to a wire lot.

Send a drawing or marked-up schematic if the same assembly contains Type S thermoelements, copper circuitry and a platinum-iridium component. Showing where each material starts and ends helps define the measurement loop and clarify which parts need thermoelectric verification.

For related design context, see medical temperature sensors and thermocouples or ultra-fine medical sensor components. To discuss a specific project, share the drawing, the requested grade and the acceptance requirements so the available scope can be reviewed.